(1) Unscrew the plastic sleeve at the front end of the unipolar surgical electrode, unscrew the metal sleeve in it, then insert the selected therapeutic contact, and then tighten the metal sleeve and the plastic sleeve in turn.

(2) Insert the therapeutic refill into the output jack of the machine.

(3) Connect a 220 volt power supply.

(4) Turn on the power switch, the indicator lights up, and hear the rotation sound of the cooling fan in the machine.

(5) Turn the adjusting knob, and you can see that the range indicated by the voltmeter is 5- 15V. If the voltage indicator is large, the output power will be large, otherwise it will be small.

(6) According to the treatment needs, turn the toggle switch to the "long fire" or "short fire" position.

(7) According to the treatment needs, when the pedal switch is needed, insert the "pedal" jack. When the foot switch is not pressed, the voltmeter has no indication and the contact has no output; When the foot switch is pressed, the voltmeter has an indication and the contact has an output, thus controlling the output. When the foot switch control is not needed, pull out the foot plug and the voltmeter will indicate. When it comes into contact with human tissue, there will be output, and when it is far away, there will be no output, so that the output can be controlled by hand.

(8) The first operator of unipolar ion therapy should hold a piece of fresh pigskin, potatoes, apples and other things with appropriate conductive moisture (if the test product does not touch the human body, the output will be very weak, and if it is not in close contact, it will spark and sting in the loose part), then approach or insert the contact, observe the corresponding relationship between the output intensity and the voltmeter indication, realize the characteristics of long fire and short fire, and practice the treatment method, and then treat the patient after being skilled in technology.

(9) During the treatment, when there is a pollution layer on the treatment contact, it will affect the output strength and need to be removed in time: the contact can be burned by hitting a metal object held by a human hand with fire or scraped off with a knife.

(10) When treating bromhidrosis or cuts, pay attention to check the therapeutic needle first. If the needle becomes dull and round due to melting, it needs to be cut off. If there is a coating on the needle body, it needs to be scraped off with a knife. During the treatment, the needle should be inserted conveniently. If it is inserted into the body too quickly, it will be due to lack.

Gap, no spark process, resulting in extremely weak destructive power.

(1 1) No matter whether it is a long fire or a short fire, the unipolar surgical electrode should be close to the human tissue, not close to it, and leave a proper gap. Only by skillfully using this gap can we give full play to the efficiency of the machine.

(12) According to the needs of disease treatment, select different unipolar electrocoagulation desalination therapeutic heads and insert them into one of the unipolar electrocoagulation jacks on the machine. The unipolar clip is inserted into another hole.

(13) Turn the toggle switch to the "long fire" position (strong curing position) or "short fire" position (weak curing position) as required.

(14) The unipolar electrocoagulation therapy needs to be used in conjunction with the pedal switch, and the "pedal" jack should be inserted first. During the treatment, all the cosmetic contacts are in contact with the tissue, and then the foot switch is pressed. When the tissue turns white, you can loosen your feet and stop the treatment. With the increase of treatment time, the solidified layer deepens, generally reaching the depth of 1 ~ 5 mm, and the treatment can be completed in 3 ~ 9 seconds.

(15) The first operator of cosmetic coagulation therapy should first take a fresh section of pig intestines (or chicken and duck intestines) for an experiment, and experience the characteristics of long fire stall (that is, strong coagulation stall) and short fire stall (that is, weak coagulation stall), as well as the output of various power levels (marked as 5 ~ 15V in the table). The solidification range is different with different solidification time.