1. Insert the power connector on the power cord into the power input socket of the equipment on the electrotherapy machine, and insert the plug into the fixed power socket.
2. Place the electrode plate on the affected part of the patient. A dry towel should be used as a cushion between the electrode and the human body, and its thickness depends on the treatment site, generally 1-2cm. The bigger the cushion, the better the thermal effect of the deep tissue. When in use, the affected part should be between the poles as far as possible. In order to prevent the electrode from shifting, the electrode and gasket can be tied to the affected area with a dry bandage.
3. According to the treatment needs, select "treatment time" (fixed time).
4. Turn "Output Selection" to "Preheat" position.
5. Press the "power switch" from "off" to "on", the green light of the "switch" is on, the power supply in the machine is turned on, the electron tube is lit, and "preheating" is started.
6. Adjust the "power supply voltage adjustment" to 220 volts indicated by the voltmeter. "Warm up" for two minutes.
7. Transfer the "Output Selection" to the "Treatment 1" file. At this time, the yellow light is on and the machine starts to work normally. The ammeter indicates the current of the oscillating tube screen.
8. Turn the "output knob" and the indication of the ammeter will change accordingly. When the resonance point is reached, the indication of the ammeter will be maximum. Treatment is best performed at the resonance point.
9. Different current values can be selected according to the needs of treatment. If the "1" file of "output selection" is not enough, you can choose "2", "3" and "4" files to increase the dose. Re-tune after each shift to make it resonate.
10. When the treatment time is up, the output will be cut off automatically, and the ammeter indicates zero. Return the "output selection" to the "preheating" position in turn, turn off the power supply, and then take out the electrode plate from the affected part of the patient.
1 1. If this machine needs to be used continuously, just turn the "output selection" to the "preheating" position when changing patients, and the yellow light will not light up and the ammeter will not indicate, so the patient can be changed. After placing the electrodes, repeat the above steps.
12. This machine is equipped with a timer, which has the functions of "normally open" and "timing". There is no need for "timing", and "treatment time" can be placed in "always open". It is necessary to automatically control the "treatment time", put the "treatment time" in the required position every time it is used, and then automatically cut off the output, which means that the machine is equivalent to returning to the "preheating" state. After changing patients, just rotate the "treatment time" and the machine will resume its output.